FDA-Inspected Site

Surgical Instruments Belgium SA (SIBEL S.A.)

6 FDA inspections are on record for Surgical Instruments Belgium SA (SIBEL S.A.) in Gembloux, Belgium.

483· Form 483 issued

Gembloux, Belgium · Last updated July 9, 2026

Overview

  • 6 FDA inspections are on record.
  • 4 inspections produced a Form 483.
  • The most recent recorded inspection date is 2026-03-19.

At a glance

Inspections

6

Form 483s

4

EIRs

0

Last outcome

Form 483

Inspection history

FDA inspections on record, most recent first.

InspectionCategoryOutcomeForm 483
16 Mar 2026 – 19 Mar 2026Medical Devices & Rad Health483View 483 →
7 Nov 2022 – 10 Nov 2022Medical Devices & Rad Health483View 483 →
27 Mar 2017 – 30 Mar 2017Medical Devices & Rad Health483View 483 →
27 Jan 2014 – 30 Jan 2014Medical Devices & Rad Health483View 483 →
15 Jan 2009Not recordedClean
15 Mar 2007Not recordedClean

Observations on record

Form 483 observations cited at this site, most recent first.

19 Mar 2026
ISO 13485:2016 Clause 7.1

Records of risk management activities in product realization not maintained

Cited by Stephen D Eich

19 Mar 2026
ISO 13485:2016 Clause 7.5.6

Planned results not consistently met during validation of processes

Cited by Stephen D Eich

19 Mar 2026
ISO 13485:2016 Clause 8.5.2

Corrective action not taken

Cited by Stephen D Eich

10 Nov 2022
21 CFR 820.100(b)

Documentation

Cited by Melanie W Pishnery

10 Nov 2022
21 CFR 820.184

DHR - not or inadequately maintained

Cited by Melanie W Pishnery

Investigators

FDA investigators tied to this site in the record.

Recurring citations

CFR citation patterns observed across this site's inspections.

CitationWhat it coversCited
21 CFR 820.90(a)Cited across public-safe Form 483 observations2x
21 CFR 820.100(a)Cited across public-safe Form 483 observations1x
21 CFR 820.100(b)Cited across public-safe Form 483 observations1x
21 CFR 820.184Cited across public-safe Form 483 observations1x
21 CFR 820.198(c)Cited across public-safe Form 483 observations1x
21 CFR 820.250(b)Cited across public-safe Form 483 observations1x
21 CFR 820.40Cited across public-safe Form 483 observations1x
21 CFR 820.50(b)Cited across public-safe Form 483 observations1x
21 CFR 820.70(a)Cited across public-safe Form 483 observations1x
21 CFR 820.70(i)Cited across public-safe Form 483 observations1x
21 CFR 820.75(a)Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.1Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.5.6Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 8.5.2Cited across public-safe Form 483 observations1x

Full site report

Get the verbatim findings behind Surgical Instruments Belgium SA (SIBEL S.A.)'s 4 Form 483s