FDA-Inspected Site

Rebuilder Medical, Inc.

9 FDA inspections are on record for Rebuilder Medical, Inc. in Charles Town, United States.

483· Form 483 issued

Charles Town, United States · Last updated July 10, 2026

Overview

  • 9 FDA inspections are on record.
  • 4 inspections produced a Form 483.
  • The most recent recorded inspection date is 2022-11-29.

At a glance

Inspections

9

Form 483s

4

EIRs

0

Last outcome

Form 483

Inspection history

FDA inspections on record, most recent first.

InspectionCategoryOutcomeForm 483
31 Oct 2022 – 29 Nov 2022Medical Devices & Rad Health483View 483 →
21 Jun 2022Foods & CosmeticsClean
20 Aug 2019 – 21 Aug 2019Foods & CosmeticsWL
20 Aug 2018 – 22 Aug 2018Medical Devices & Rad Health483View 483 →
20 Oct 2014Medical Devices & Rad HealthClean
8 Jul 2011Not recordedClean
11 Jan 2011Not recordedWLView 483 →
17 Aug 2009Not recorded483View 483 →
28 Oct 2008Not recordedClean

Observations on record

Form 483 observations cited at this site, most recent first.

29 Nov 2022
21 CFR 820.100(b)

Documentation

Cited by Dennis R Hock, Nadia R Barreda

29 Nov 2022
21 CFR 820.181(a)

DMR device specifications

Cited by Dennis R Hock, Nadia R Barreda

29 Nov 2022
21 CFR 820.198(c)

Investigation of device failures

Cited by Dennis R Hock, Nadia R Barreda

29 Nov 2022
21 CFR 820.20(b)

Lack of or inadequate organizational structure

Cited by Dennis R Hock, Nadia R Barreda

29 Nov 2022
21 CFR 820.30(g)

Design Validation - Risk analysis not performed/inadequate

Cited by Dennis R Hock, Nadia R Barreda

Investigators

FDA investigators tied to this site in the record.

Recurring citations

CFR citation patterns observed across this site's inspections.

CitationWhat it coversCited
21 CFR 820.198(a)Cited across public-safe Form 483 observations4x
21 CFR 820.100(a)Cited across public-safe Form 483 observations2x
21 CFR 820.30(g)Cited across public-safe Form 483 observations2x
21 CFR 820.90(a)Cited across public-safe Form 483 observations2x
21 CFR 820.100(b)Cited across public-safe Form 483 observations1x
21 CFR 820.181(a)Cited across public-safe Form 483 observations1x
21 CFR 820.181(b)Cited across public-safe Form 483 observations1x
21 CFR 820.184Cited across public-safe Form 483 observations1x
21 CFR 820.198(c)Cited across public-safe Form 483 observations1x
21 CFR 820.198(e)Cited across public-safe Form 483 observations1x
21 CFR 820.20(b)Cited across public-safe Form 483 observations1x
21 CFR 820.20(c)Cited across public-safe Form 483 observations1x
21 CFR 820.22Cited across public-safe Form 483 observations1x
21 CFR 820.25(b)Cited across public-safe Form 483 observations1x
21 CFR 820.30(a)Cited across public-safe Form 483 observations1x
21 CFR 820.50Cited across public-safe Form 483 observations1x
21 CFR 820.80(d)Cited across public-safe Form 483 observations1x
21 CFR 820.80(d)(3)Cited across public-safe Form 483 observations1x
21 CFR 820.90(b)(2)Cited across public-safe Form 483 observations1x

Full site report

Get the verbatim findings behind Rebuilder Medical, Inc.'s 4 Form 483s

Source PDFs, verbatim observations, and recurring CFR citations, so your team knows what this site gets cited for.

  • The site's full inspection history, with every Form 483 and its outcome
  • Warning letters and recalls tied to the company behind it
  • Whether recurring CFR citations get closed out or repeat