Form 483

Medical Informatics Corp. Form 483, 2026-03-27

FDA issued a Form 483 to Medical Informatics Corp. on 2026-03-27 with 7 public-safe observations.

483· Form 483 issuedMedical Devices & Rad Health7 observations

Medical Informatics Corp. · Houston, United States · Issued March 27, 2026

Overview

  • 7 public-safe observations are available for this inspection.
  • 7 structured CFR citations are represented.
  • 1 investigator is tied to this Form 483.

Inspectional observations

Each observation cites the regulation FDA found the firm out of compliance with.

27 Mar 2026
21 CFR 803.17

Lack of Written MDR Procedures

Cited by Jocelyn C Turner

27 Mar 2026
ISO 13485:2016 Clause 4.1.2

Risk-based approach not applied to QMS processes

Cited by Jocelyn C Turner

27 Mar 2026
ISO 13485:2016 Clause 5.6.3

Management review output records lack input reviewed, decisions, and actions

Cited by Jocelyn C Turner

27 Mar 2026
ISO 13485:2016 Clause 7.3.1

Design and development procedures not documented

Cited by Jocelyn C Turner

27 Mar 2026
ISO 13485:2016 Clause 7.4.1

Supplier monitoring and re-evaluation not planned

Cited by Jocelyn C Turner

Inspected site

MIMedical Informatics Corp.Houston, United States

Inspection team

Cited regulations

CFR citations associated with this Form 483.

CitationWhat it coversCited
21 CFR 803.17Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 4.1.2Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 5.6.3Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.3.1Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.4.1Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.5.6Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 8.5.2Cited across public-safe Form 483 observations1x

Full 483 report

Get the verbatim text of all 7 observations

The full inspection record, as Atlas holds it, so your team sees exactly what was cited and how it was answered.

  • The site's full inspection history around this Form 483
  • Whether these citations escalated to a warning letter
  • The investigators involved and the citations they repeat elsewhere