Inspectional observations
Each observation cites the regulation FDA found the firm out of compliance with.
28 Apr 2026
ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
Cited by Jocelyn C Turner
28 Apr 2026
ISO 13485:2016 Clause 7.3.4
Records of design and development outputs not maintained
Cited by Jocelyn C Turner
28 Apr 2026
ISO 13485:2016 Clause 7.3.6
Records of design and development verification not maintained
Cited by Jocelyn C Turner
28 Apr 2026
ISO 13485:2016 Clause 7.3.9
Requirements for design and development changes not met before implementation
Cited by Jocelyn C Turner
Inspected site
LSLion Street Medical, dba Pensar Medical LLCHouston, United States
Inspection team
Cited regulations
CFR citations associated with this Form 483.
| Citation | What it covers | Cited |
|---|---|---|
| 21 CFR 803.17(a)(1) | Cited across public-safe Form 483 observations | 1x |
| ISO 13485:2016 Clause 4.1.2 | Cited across public-safe Form 483 observations | 1x |
| ISO 13485:2016 Clause 7.3.4 | Cited across public-safe Form 483 observations | 1x |
| ISO 13485:2016 Clause 7.3.6 | Cited across public-safe Form 483 observations | 1x |
| ISO 13485:2016 Clause 7.3.9 | Cited across public-safe Form 483 observations | 1x |
| ISO 13485:2016 Clause 8.2.2 | Cited across public-safe Form 483 observations | 1x |
| ISO 13485:2016 Clause 8.5.2 | Cited across public-safe Form 483 observations | 1x |
Full 483 report
Get the verbatim text of all 7 observations
The full inspection record, as Atlas holds it, so your team sees exactly what was cited and how it was answered.
- The site's full inspection history around this Form 483
- Whether these citations escalated to a warning letter
- The investigators involved and the citations they repeat elsewhere