Form 483

Jiangsu Suyun Medical Materials Co., Ltd. Form 483, 2026-04-10

FDA issued a Form 483 to Jiangsu Suyun Medical Materials Co., Ltd. on 2026-04-10 with 4 public-safe observations.

483· Form 483 issuedMedical Devices & Rad Health4 observations

Jiangsu Suyun Medical Materials Co., Ltd. · Lianyungang, China · Issued April 10, 2026

Overview

  • 4 public-safe observations are available for this inspection.
  • 4 structured CFR citations are represented.
  • 1 investigator is tied to this Form 483.

Inspectional observations

Each observation cites the regulation FDA found the firm out of compliance with.

10 Apr 2026
ISO 13485:2016 Clause 7.3.6

Design and development verification not performed

Cited by Maryam Tabatabaie

10 Apr 2026
ISO 13485:2016 Clause 7.3.9

Procedures to control design and development changes not documented

Cited by Maryam Tabatabaie

10 Apr 2026
ISO 13485:2016 Clause 7.6(b)

Adjustment to measuring equipment was not performed or recorded

Cited by Maryam Tabatabaie

10 Apr 2026
ISO 13485:2016 Clause 8.4

Procedures for analysis of data not documented

Cited by Maryam Tabatabaie

Inspected site

Inspection team

Cited regulations

CFR citations associated with this Form 483.

CitationWhat it coversCited
ISO 13485:2016 Clause 7.3.6Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.3.9Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 7.6(b)Cited across public-safe Form 483 observations1x
ISO 13485:2016 Clause 8.4Cited across public-safe Form 483 observations1x

Full 483 report

Get the verbatim text of all 4 observations

The full inspection record, as Atlas holds it, so your team sees exactly what was cited and how it was answered.

  • The site's full inspection history around this Form 483
  • Whether these citations escalated to a warning letter
  • The investigators involved and the citations they repeat elsewhere